First five minutes
Stop treating. Power the system down. Keep water away from the floor connectors and the wall outlet, and do not power back up to 'check if it still leaks'. Note where the water appeared — under the chassis, at the handpiece, at a rear fitting — because location is half the diagnosis. Photograph it.
Where leaks come from
- Aged or perished tubing and O-rings inside the cooling loop
- A cracked reservoir or a failing pump seal
- Fittings disturbed when the machine was moved
- A handpiece cable with internal coolant lines (diode platforms) damaged at the flex point
- Condensation mistaken for a leak — real, but a technician should be the one to call it
Why it cannot wait
Coolant loops run next to high-voltage power supplies. A slow leak corrodes boards and connectors quietly for weeks before the dramatic failure, and it changes a fitting-and-tubing job into a board job. Early leaks are cheap; late leaks are not.
What we do
We pressure-test the loop, trace the source, replace the failed seals, tubing or components, refill with the correct fluid, and run the system long enough to prove the repair dry — then check the electronics the water visited.
FAQ
Questions on this topic
It is only a few drops. Can I finish the day?
We advise against it. You cannot see where the rest of the water went, and the parts at risk are the expensive ones.
Could it just be condensation?
Possibly, in a humid room with a cold loop — but the safe move is the same: stop, note, and have it checked before the next clinic day.